STATIM5000问题解答-循环错误 循环错误序号问题描述可能的原因 CycleFault#1在规定的时间内卡 式盒温度无法达到 95℃ 可能由于灭菌盒装载过多或者盒体泄漏引起了热保险丝的 熔断。热保险丝熔断有可能是因为水泵输出过弱和/或蒸汽 发生器被污染或控制蒸汽发生器的三端可控硅损坏引起。如 果三端可控硅损坏了,则热保险...
STATIM 5000 问题解答-循环错误.pdf,STATIM 5000 问题解答-循环错误 循环错误序号 问题描述 可能的原因 Cycle Fault #1 在规定的时间内卡 可能由于灭菌盒装载过多或者盒体泄漏引起了热保险丝的 式盒温度无法达到 熔断。热保险丝熔断有可能是因为水泵输出过弱和/或蒸汽 95
Safetycompliance5CycleFaultNumbersanddescriptions forrevision3.xand4.xcontrollers66 1.IdentifyingSTATim2000Units7CycleFaultNumbersanddescriptionsonunits Identifyingsoftwareversions7withrevision2.x/5.x/6.xand7.xtypecontroller Identifyingcontrollerboardtypes9boards(1.xx/2.xx/4.xx/5.xxand6.xxsoftware67 ...
Statim autoclaves can all self-diagnose any fault in the sterilization cycle for easy maintenance in the field, and most can be corrected without a technician. How to safely maintain a STATIM Regular operator maintenance Refer to the latest Operator Manual for regular cleaning, seal and filter r...