If employing plastic apparatus, such as microplates and pipet tips for auto- matic pipetters, use apparatus that is shown to be free of detectable endotoxin and does not interfere in the test. [NOTE—In this chapter, the term “tube” includes any other receptacle such as a microtiter well...
USP/NF <161> and <85> Ph Eur 2.6.14. Method A, C, and D LAL Testing Panel Kinetic Chromogenic:Using the kinetic chromogenic test, a sample is mixed with lysate. In the presence of endotoxins, the lysate reacts with the endotoxin, resulting in a chromophore release from a chromogenic ...
endotoxin and does not interfere in the test NOTE In this chapter the term tube includes any other receptacle such as a microtiter well REAGENTS AND TEST SOLUTIONS Amoebocyte Lysate A lyophilized product obtained from the lysate of amoebocytes white blood cells from the horseshoe crab Limulus poly...
Accessed from 10.6.1.1 by belmont7 on Tue Nov 01 10:28:09 EDT 2016 USP 40 Biological Tests / á85? Bacterial Endotoxins Test 165 DETERMINATION OF MAXIMUM VALID DILUTION (MVD) The maximum valid dilution is the maximum allowable dilution of a specimen at which the endotoxin limit can be ...
Units, U S P EndotoxinUspeu, OneReagents, Since L A LUSP (2003) Bacterial Endotoxins Test. USP 26 NF 21, 2023–2026USP24/NF19,Bacterial endotoxins test[S].Bacterial Endotoxins Test.Bacterial Endotoxins Test.USP 23 & NF18. 1995Bacterial endotoxins test. USPC,Inc. . 2000...
USP 161 USP 797 AAMI ST72 EP 2.6.14 JP 4.01 Testing Options Ask an expert for a specific consultation on your product. If you are ready to submit your samples for testing, fill out theSample Submission Form. LAL110 Add to QuoteBacterial Endotoxin Test: Kinetic turbidimetric method ...
. If employing plastic apparatus, such as microplates and pipet tips for automatic pipetters, use only that which has been shown to be free of detectable endotoxin and not to interfere with the test.[note—In this chapter, the term “tube” includes any other receptacle such as a micro-tit...
USP Chapter <85> first appeared as an informational chapter in USP XX/NF XV (1980) describing an alternative test to the RPT. The methods described in <85> are not tests for all pyrogens. The BET detects only endotoxin pyrogens from Gr...
If employing plastic apparatus, such as microplates and pipet tips for automatic pipetters, use apparatus that is shown to be free of detectable endotoxin and does not interfere in the test. [NOTE—In this chapter, the term "tube" includes any other receptacle such as a microtiter well.] ...
limits. However, if Wfisamples are being tested to an endotoxin limit of NMT 0.25 EU/mL, a less sensitive assay (e.g. with LOD of 0.125 EU/mL) may be used which will dramatically reduce the assay runtime (e.g. to 25 minutes as shown above for the kinetic turbidimetric test). ...